Product Center

Taxo™ P Discs
Cat. # Desc. Qty. Unit
231554 Taxo™ P Discs 10 SP
Intended Use:

Taxo™ P Discs are recommended for use in the presumptive identification of pneumococci, and are intended for use with pure cultures.

Catalog # Description Code Quantity Unit
231554 Taxo™ P Discs
For differentiating Pneumococci
P 10 SP

BBL™ Taxo™ Discs for Differentiation of Pneumococci
Revision: 0604    Revision Date: 06/01/2004   

Intended Use: «

Taxo™ P Discs are recommended for use in the presumptive identification of pneumococci, and are intended for use with pure cultures.

Intended Use: » Intended Use:

Taxo™ P Discs are recommended for use in the presumptive identification of pneumococci, and are intended for use with pure cultures.


Product Summary: «

Taxo P Discs are impregnated with ethylhydrocupreine hydrochloride (optochin), a drug used for pneumonia therapy before sulfonamides became available. The growth of pneumococci, but not of other streptococci, is markedly inhibited by this chemical.1,2 Pneumococci may, therefore, be differentiated from other alpha-hemolytic streptococci by the formation of a zone of inhibition around a Taxo P disc placed on a blood agar plate heavily inoculated with a pure culture suspected to be Streptococcus pneumoniae.2

Product Summary: » Product Summary:

Taxo P Discs are impregnated with ethylhydrocupreine hydrochloride (optochin), a drug used for pneumonia therapy before sulfonamides became available. The growth of pneumococci, but not of other streptococci, is markedly inhibited by this chemical.1,2 Pneumococci may, therefore, be differentiated from other alpha-hemolytic streptococci by the formation of a zone of inhibition around a Taxo P disc placed on a blood agar plate heavily inoculated with a pure culture suspected to be Streptococcus pneumoniae.2


User Quality Control: « User Quality Control:

At the time of use, check performance with pure cultures of stable control organisms producing known desired reactions. The use of S. pneumoniae ATCC™ 6305 is recommended to demonstrate zone formation (14 mm or larger in diameter) and S. pyogenes ATCC 12384 is recommended for use as a negative (no zone) control.

Quality control requirements must be performed is accordance with applicable local, state and/or federal regulations or accreditation requirements and your laboratory's standard Quality Control procedures. It is recommended that the user refer to pertinent NCCLS guidance and CLIA regulations for appropriate Quality Control practices.


Reagents: « Reagents:

Taxo P Discs are 6 mm discs made from high quality absorbent paper impregnated with approximately 5.0 µg of ethylhydrocupreine hydrochloride per disc. The last disc in each cartridge is marked "X" and contains the agent as coded.


Warnings and Precautions: « Warnings and Precautions:

For in vitro Diagnostic Use.

Observe aseptic techniques and established precautions against microbiological hazards throughout all procedures. After use, test plates and other contaminated materials must be sterilized by autoclaving before discarding.



Information shown on this page is a short summary extracted from the Package Insert, available as a PDF under the Related Documents section of this page.