Preanalytical Variables in Urine Testing
|
| |||||||||||||||||||||||||||||||||||
There are many preanalytical variables that can introduce error into laboratory test results. When unexpected results are seen in urine measurements, they should be interpreted in the context of other analytes and clinical results that measure similar aspects of renal function. An example is blood and bacteria testing in routine urinalysis. A false-positive result for blood can be obtained when a specimen contains infection-causing bacteria. The microbial peroxidase activity can cause a false-positive reaction for blood. In this case, microscopic analysis will verify the presence of bacteria and possibly the lack of red blood cells. This will confirm that the response for blood on the reagent strip is inaccurate. Another example is the effect of urine color on urinalysis reagent test pads. Abnormally colored urine due to medications, food dyes, or vitamins can alter the color change on dipstick test pads for tests other than color. This can be ruled out by doing a visual exam of the urine specimen prior to performing the dipstick. For this reason, the College of American Pathologists (CAP) recommends that laboratories have a procedure for the correlation of microscopic and macroscopic results. In general, collection time, transport, and storage conditions can be examined to determine the causes of error. There are some basic guidelines that should be followed for urinalysis and the storage and handling of urinalysis dipsticks. By following these precautions, inaccurate results can be minimized.
Typical urinalysis dipstick tests include: specific gravity, pH, protein, blood, nitrite, leukocyte esterase, glucose, ketones, bilirubin, and urobilinogen. Color and clarity can be measured either visually or by an analyzer. The following will be a review of some of the preanalytical variables that can affect these tests. Preanalytical VariablesColorThe color of urine, which is normally colorless or one of the various shades of yellow, can be altered by medications, vitamins, dyes, or diet. If an unusual color is detected for the urine specimen, one of these conditions could be the cause. Once the cause is determined, it should be noted in the laboratory results. Some abnormal urine colors and their possible causes are:
These abnormal urine colors can affect other dipstick results by causing a colorimetric reaction that may be misinterpreted by the instrument and give incorrect results. It is important that the tube be of a clear material when determining the color of the urine. Using good lighting and a white background helps to ensure the color is being read accurately and consistently. Color descriptions should be standardized. ClarityAnother visual measurement is clarity. (A normal urine specimen is typically clear.) Urine clarity can be related to the handling conditions of the specimen. If a urine specimen is old and unpreserved, it can become cloudy from bacterial overgrowth. In turn, if a specimen has been stored in a refrigerator, amorphous urates or phosphates can cause temporary cloudiness (which dissolve when the specimen is brought to room temperature). Amorphous urates are observed in acidic urines and amorphous phosphates are found in alkaline urine. The collection time and storage conditions of the specimen should be reviewed to determine if cloudiness may be caused by storage conditions. Other causes of cloudy samples include talcum powder, mucus, crystals, leukocytes, epithelial cells, and fat. Clear tubes are best for examining urine clarity. Result reporting should be standardized, with clear, hazy, cloudy, and turbid being the most commonly used descriptors. Specific gravity, or the amount of dissolved particles in a solution, is another measurement performed on urine. Specific gravity is affected by the number, amount, and weight of solutes in the specimen. It serves as a measure of the kidney’s ability to dilute and concentrate urine. A normal random urine has a specific gravity range of 1.001–1.035. Dehydration, sweating, diarrhea, radiopaque dyes, and antibiotics can cause high results because the ratio of dissolved particles in low volumes of solute will be elevated. A high fluid intake or consumption of diuretics can cause low measurements because of the low quantity of dissolved particles in a large volume of solute. The pH test indicates whether a specimen is acidic (pH <7) or alkaline (pH >7). A normal urine pH ranges from 5-7 and is a useful tool for the laboratorian, often predicting what may be seen in subsequent microscopic examination. Certain crystals exist in either an acidic or alkaline environment. Some examples of these are uric acid or calcium oxalate crystals in acidic urine and calcium carbonate or magnesium phosphate in alkaline urine. Dilute and alkaline urines can dissolve casts and cells. Bacterial overgrowth in a specimen standing at room temperature will produce a higher pH due to the conversion of urea into ammonia.1 Diets high in vegetables, citrus fruits, and dairy produce an alkaline pH. Lower pH levels may be seen in uncontrolled diabetes or may reflect a diet high in meat or cranberries. Starvation and diarrhea can produce a more acidic urine. Lastly, mishandling the reagent strips by allowing runover from the protein reagent pad can cause false negative pH results. Protein is a very important analyte measured in urine because it monitors the function of the kidneys. A normal urine specimen should not contain more than trace amounts of protein. Most reagent pads primarily measure albumin and, to a lesser extent, other proteins. Residue of disinfectants in contaminated urine containers can cause false-positive protein results. Similar false positives may be seen after strenuous exercise and in highly alkaline urines. Urine specimens with a high specific gravity can produce trace results. Dilute specimens, fever, mental stress, mucus, and exposure to extreme heat or cold can produce falsenegative protein results. Blood is another measurement on the urine dipstick, detecting intact red blood cells, free hemoglobin, and myoglobin. A normal urine specimen is usually negative for blood. False positives can be caused by the presence of chlorine bleach, consumption of colored medications by the patient, or a microbial peroxidase reaction from bacterial presence. False-positive-causing contaminants can also be introduced during time of collection, such as in specimens submitted by women during menstruation. High levels of a reducing substance such as ascorbic acid at >25 mg/dL can produce falsenegative results. False-negative blood results can also occur when formalin is used as a preservative, when there is>10 mg/dL of nitrite, or when mixing is inadequate and red blood cells have settled in the tube. Nitrite is also a test on urine dipsticks. A normal urine specimen is negative for nitrite. Some of the ways in which falsepositive results can occur include bacterial overgrowth in specimens that have not been stored properly, colored medications, and dyes. False-negative results can be seen with high levels of ascorbic acid, >25 mg/dL, or when dietary nitrate is missing due to starvation, IV feeding, or fasting. The conversion of nitrate to nitrite takes about 4 hours in the bladder. If this hasn’t occurred, results could be negative for the presence of nitrite. This can be seen in randomly collected specimens. Leukocyte esterase is an indicator of white blood cells. It is typically negative in normal specimens. False positives can occur when collection containers have been contaminated with chlorine bleach or other oxidizing detergents. Other influences that can produce false-positive results include formalin as a preservative and vaginal discharge. High specific gravity, some antibiotics, such as tetracycline, and large amounts of glucose or ascorbic acid can cause false negative leukocyte esterase results. Glucose, another urine dipstick test, is mainly used to monitor diabetes. Normal urines are negative for glucose. Some normal specimens have small amounts of glucose, that are below levels of sensitivity for the reagent strip. Just like leukocyte esterase, false-positive glucose results are achieved when collection devices have been exposed to chlorine bleach or detergents. Improper storage of reagent strips, when exposed to air, have been noted to produce false-positive results. Ascorbic acid, >50 mg/dL, is again a culprit in causing false-negative glucose results. Over time at room temperature, glucose will decrease due to glycolysis from bacteria. Tetracycline has been determined to cause false-negative glucose results, and refrigerated specimens that were not allowed to reach room temperature can produce false-negative results because the enzymatic reaction is affected. Ketone bodies are a by-product of fat breakdown. Normal urines are negative for ketones. Increased ketones can be due to starvation or alcoholism. Also strenuous exercise, fever, fasting, vomiting, and high protein diets can cause high ketone values. Urine specimens with a high specific gravity and low pH have been known to cause trace ketone results. Acetone, one by-product of fat breakdown, evaporates rapidly if the uncovered specimen is left standing at room temperature. Specimens should, therefore, be tested immediately for ketones or refrigerated in a closed container.2 Some urinalysis preservatives can maintain ketone levels. It is important to check the manufacturer’s claims. Bilirubin measures liver function and is usually negative in normal urine specimens. Colored medications, those in the yellow, orange, and red spectrum, can cause false-positive results. Ascorbic acid, >25 mg/dL, causes false negative results, as can specimens with high nitrite concentrations. Exposure to light causes bilirubin to degrade over time, leading to false-negative results. It is recommended that specimens submitted for bilirubin measurements be kept in a dark place or collected in an amber container. Urobilinogen is another measurement of liver function. A normal urobilinogen result is <1 Ehrlich unit/dL. The same factors that affect bilirubin (exposure to light and colored medications) also affect urobilinogen. In addition, formalin, which is sometimes used as a preservative, will cause false-negative urobilinogen results, as can ascorbic acid or nitrite. Standardization when processing urine specimens for microscopic sediment analysis has become a very important guideline as recommended by the Clinical and Laboratory Standards Institute (CLSI, formerly known as NCCLS). This includes the use of engineered tubes, pipettes, and standardized calibrated slides. The best types of tubes for microscopic sediment analysis are clear, plastic tubes with conical bottoms. A cap or lid and volume gradations are valuable features. CLSI does not support the use of glass slides and cover slips due to the lack of sample volume standardization. The chambers on the specially designed slides are calibrated for a specific urine sediment volume that ensures standardization.
As part of specimen processing standardization, the ratio of the volume of urine required to fill the tube is compared with the volume of urine and sediment, which remain in the tube after centrifugation and decanting. Ensuring that these two are maintained is one of the keys to standardizing microscopic sediment analysis. CLSI recommends centrifugation of urine specimens at 400 RCF (g). Most hospitals typically centrifuge for 5 minutes. Any variations in speed and time can change the cellular elements obtained in the sediment. All equipment used for sediment processing and analysis should be properly maintained. Centrifuges should be calibrated and microscopes should be quality-control and proficiency tested regularly to guarantee accurate testing. Unpreserved specimens that have been unrefrigerated for more than 2 hours from time of collection should not be accepted for microscopic analysis due to the increase in bacterial overgrowth and the disintegration of cells and casts. The urine becomes alkaline, causing red blood cells and white blood cells to lyse and casts to dissolve. If a specimen has been refrigerated for storage, it should be allowed to come to room temperature and mixed well prior to analysis. Amorphous urates or phosphates develop in cold conditions and will affect the analysis. Contaminants that can be seen during sediment analysis include mucus, spermatozoa, fibers, talcum powder, and oil. It is important not to confuse these contaminants with cellular components. Stains can help in the identification of cells. Flow cytometry is another method of examining the urine for microscopic elements. This method of microscopic analysis, however, is almost unaffected by preanalytical variables. The greatest factor that could affect flow cytometry test results is insufficient mixing of the specimen. Urine Culture and SensitivityThe microbiology lab also conducts clinical testing on urine. Preanalytical variables that could affect culture and sensitivity testing include:
Urine ChemistryRandom testing of urine chemistry analytes is not usually considered valuable because of the lack of collection of these analytes at any given time in the bladder. Timed specimens provide the most valuable information for the concentration of a specific analyte. Some of the preanalytical variables that affect these tests are redundant to those for urine dipstick chemistries. There are various factors, that can affect any one of these analytes. The most common influences are preservatives, diet, and medication. More specifically some of the drugs and foods that affect urine chemistry results are as follows:
It is recommended that any drug known to interfere with a test be discontinued until a drug-free specimen can be produced. It is important that the total urine volume be collected and recorded during the time period in order to correctly calculate the analyte concentrations. Other guidelines that should be followed when performing urine chemistry testing are: Consideration should be given as to whether preserved or unpreserved specimens are used. Proliferation of bacteria could be a concern with unpreserved specimens. Potential interference with assay methods should be considered when chemical preservatives are used. Each institution should perform evaluations to make these determinations for their facility based on their populations and levels of clinical acceptance. NCCLS GP16-A2 provides a table of common 24-hour urine preservatives corresponding to the chemistry tests. The most commonly cited methods for preservation were refrigeration, freezing, HCl acid, boric acid, and acetic acid.3 Specimens may need to be split if various tests that require different preservatives are requested. This can be accomplished by pooling the collections over time, then evenly distributing the specimen into multiple containers. Another option would be to collect multiple 24-hour specimens. Drugs of abuse are tested on urine. The preanalytical variables that affect testing drugs of abuse in urine are listed below:
Urine is a valuable specimen for the screening, diagnosis, and monitoring of diseases. It can also be used to monitor therapy. There are various collection methods for acquiring this specimen, each of which has its merits. By minimizing the preanalytical influences during specimen collection and handling, a suitable specimen can be obtained. Specimen handling is probably the most important step in obtaining a good urine specimen that can provide the most useful clinical information. It is also important to carefully record any details that may be beneficial in the interpretation of the urine results. After all, the results are only as good as the specimen collected. References
|
For questions regarding our Blood, Urine collection, or other sample collection products and services please contact us.