BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Citrate

Performance Characteristics

(Using 4 mL Draw Capacity)

Table 2 shows the reproducibility of separated sample quality using the BD Vacutainer® CPT™ Tube system which was tested and compared to the FICOLL™ HYPAQUE™ method. Ten samples of one donor's blood were centrifuged and assayed in duplicate for each method. No final washing steps were performed. The samples were resuspended to approximately equal final volumes. Between Tube variation was calculated by differencing the mean of the duplicate readings for each Tube.

Table 2

Reproducibility Study: Recovery, Purity, Viability,Red Blood Cell (RBC) and Granulocyte (PMN) Contamination** for BD Vacutainer® CPT™ (CPT) Citrate Tube Method and FICOLL™ HYPAQUE™ (FH) Method
PARAMETER   Number MEAN**** SD CV%
Recovery* CPT
FH
9***
10
70.3
82.6
9.7
5.5
13.8
6.7
Purity*          
Total Mononuclear CPT 10 98.3 0.6 0.6
Cells FH 10 99 0.3 0.3
Lymphocytes CPT
FH
10
10
84.4
84.8
1.3
0.9
1.5
1.1
Monocytes CPT
FH
10
10
13.9
14.2
1.4
0.8
10.4
5.9
Viability CPT
FH
10
10
100
99.8
0.0
0.4
0.0
0.4
RBC Contamination CPT
FH
10
10
7.4
2.6
1.6
1.0
21.7
39.0
PMN Contamination CPT
FH
10
10
1.7
1.0
0.6
0.3
34.5
30.7
Footnotes:
****Recovery and Purity are based on the Mean of 2 readings.
****Parameters defined in Table 1
****Sample lost due to spillage, Recovery could not be calculated.
****Using an F-Statistic (2-tailed) at the 5% level, no significant difference was detected between the methods.
• MEAN - Arithmetic Mean
• SD - Standard Deviation
• CV - Coefficient of Variation (%)
• N - Number of Tubes

Table 3 shows the cell percentage obtained from 4 groups of 5 patients each, using the FICOLL™ HYPAQUE™ and the BD Vacutainer® CPT™ Tube cell separation methods. Recovery and purity percentage were determined from the mean of duplicate values obtained using the Coulter® STKR cell counting method. Viability percentages were determined by Acridine Orange/Ethidium Bromide staining(7).

Table 3

Results from Patient Sample Study
SAMPLE
NUMBER
TOTAL
MONONUCLEAR
RECOVERY
(%)
PURITY
(%)
VIABILITY
(%)
  CPT FH CPT FH CPT FH
Leukemia            
1 CLL 70.5 76.2 98.0 99.4 96.4 97.4
2 CLL 67.8 70.6 99.0 98.5 99.2 98.9
3 CLL 79.7 83.2 97.6 99.5 96.0 97.1
4 CLL 71.4 68.3 99.1 99.3 97.5 98.0
5 CLL 75.6 82.4 99.7 99.7 99.4 100.0
HIV Positive            
6 HIV 62.1 67.3 98.4 97.8 98.5 98.0
7 HIV 59.0 63.7 97.4 98.3 100.0 99.5
8 HIV 68.3 65.6 99.2 96.9 99.2 100.0
9 HIV 63.9 73.0 98.2 97.9 99.3 100.0
10 HIV 70.7 74.8 99.2 98.5 97.2 99.4
Diabetics            
11 DIAB 58.1 64.3 96.7 97.0 99.5 97.8
12 DIAB 66.4 72.3 98.9 99.3 98.9 99.6
13 DIAB 70.2 67.5 97.5 99.0 100.0 99.7
14 DIAB 64.9 70.7 98.5 96.5 99.7 98.2
15 DIAB 62.7 67.7 97.0 98.2 97.8 98.3
Auto Immune            
16 SLE 62.3 64.5 96.4 95.7 97.5 99.0
17 SLE 60.5 63.4 97.0 96.3 99.2 97.3
18 SLE 65.6 62.3 95.7 96.9 96.5 99.0
19 RA 71.5 66.5 96.2 97.2 99.2 100.0
20 RA 69.5 75.6 95.7 97.0 98.9 99.2

Footnotes:
• CLL - Chronic Lymphocytic Leukemia
• HIV - Human Immunodeficiency Virus Positive
• DIAB - Type I Diabetes
• SLE - Systemic Lupus Erythematosis
• RA - Rheumatoid Arthritis
• Recovery, Purity, and Viability Parameters defined in Table 1

FICOLL is a trademark of Amersham Biosciences AB
HYPAQUE is a trademark of Amersham Health AS

About BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Citrate

Intended Use

Summary & Explanation

Test Principles

Reagents, Supplies & Equipment

Warnings & Precautions

Venipuncture Technique

Procedure

Limitations

Expected Results

Performance Characteristics

Troubleshooting

Product Insert [PDF]

References

How To Order

Contact Us

More Molecular Diagnostics & Proteomics products



 

[spacer image]
[spacer image] [spacer image] [spacer image]
[spacer image] [spacer image] [spacer image]
[spacer image]