Expected Normal Donor Study Results
(Using 4 mL Draw Capacity)
Table 1 shows the cell percentages
obtained from 45 blood specimens from 32 normal healthy adults using
the
FICOLL™ HYPAQUE™ and the BD Vacutainer® CPT™ Tube cell
separation methods. Recovery and Purity percentages were determined
from the mean of duplicate values obtained using the Coulter Counter®
Model S Plus IV. Viability percentages were determined by Acridine
Orange/Ethidium Bromide staining(7). Red blood cell percentages were
determined by hemocytometer count under a light microscope. Both viability
and red blood cell contamination were single determinations for each
specimen.
Table 1
Cell Percentages, BD Vacutainer® CPT™ (CPT) Citrate Tube Method and
FICOLL™ HYPAQUE™ (FH) Method
| PARAMETER |
|
MEAN |
SD |
CV |
| Recovery* |
CPT
FH |
71.7
80.9 |
10.5
8.5 |
14.7
10.5 |
| Purity* |
|
|
|
|
| Total Mononuclear |
CPT |
98.0 |
1.8 |
1.8 |
| Cells |
FH |
98.7 |
0.8 |
0.9 |
| Lymphocytes |
CPT
FH |
85.9
87.3 |
4.3
4.0 |
5.0
4.5 |
| Monocytes |
CPT
FH |
12.1
11.4 |
3.3
3.4 |
27.7
29.7 |
| Viability |
CPT
FH |
99.9
99.7 |
0.3
0.5 |
0.3
0.5 |
| RBC Contamination |
CPT
FH |
14.3
6.8 |
9.4
5.5 |
65.9
81.3 |
| PMN Contamination |
CPT
FH |
2.0
1.2 |
1.8
0.8 |
88.7
68.1 |
|