Performance Characteristics
Table 2 shows the repeatability
of sample preparation using the BD Vacutainer® CPT™ Tube
system which was tested and compared to the
FICOLL™ HYPAQUE™ method.
Ten samples of one donor's pooled blood were centrifuged and assayed
in duplicate for each method. No final washing steps were performed.
The samples were resuspended to equal final volumes. Estimates of
variability due to repeated measurements are shown for within tube
and between tubes. Within tube variation was measured by taking duplicate
readings from each tube.
Table 2
Repeatability of recovered Mononuclear Cell values, (FICOLL™ HYPAQUE™ versus BD Vacutainer® CPT™ Tube Method) (7)
| PARAMETER |
| % VARIATION |
| AVG |
WITHIN TUBE* |
BETWEEN TUBE |
| % |
SD |
CV |
SD |
CV |
| Recovery |
|
|
|
|
|
| BD Vacutainer® CPT™ (n=9) |
79.5 |
3.41 |
4.3 |
6.25 |
7.9 |
| FICOLL™ HYPAQUE™ (n=10) |
73.3 |
3.55 |
4.8 |
2.22 |
3.0 |
| Purity |
| BD Vacutainer® CPT™ (n=9) |
|
|
|
|
|
| Total Mononuclear Cells |
94.6 |
0.69 |
0.7 |
1.48 |
1.6 |
| Lymphocytes |
72.6 |
0.93 |
1.3 |
0.87 |
1.2 |
| Monocytes |
22.0 |
0.94 |
4.3 |
0.99 |
4.5 |
| FICOLL™ HYPAQUE™ (n=10) |
|
|
|
|
|
| Total Mononuclear Cells |
96.5 |
0.66 |
0.7 |
1.46*** |
1.5 |
| Lymphocytes |
80.5 |
0.82 |
1.0 |
0.91 |
1.1 |
| Monocytes |
16.0 |
1.06 |
6.6 |
1.30 |
8.2 |
| RBC Contamination |
| BD Vacutainer® CPT™ (n=9) |
28.4 |
1.36 |
4.8 |
13.73 |
48.5 |
| FICOLL™ HYPAQUE™ (n=10) |
14.9 |
1.21 |
8.1 |
5.85 |
39.3 |
| Granulocyte Contamination |
| BD Vacutainer® CPT™ (n=9) |
5.4 |
0.62 |
11.4 |
1.49 |
27.3 |
| FICOLL™ HYPAQUE™ |
3.5 |
0.57 |
16.2 |
1.47 |
42.1 |
|
|