Expected Normal Donor Study Results
Table 1 shows the cell percentages
obtained from forty-two blood specimens from a total of twenty-nine
normal apparently healthy adults using the
FICOLL™ HYPAQUE™ and the
BD Vacutainer® CPT™ Tube cell separation methods. Recovery
and Purity percentages were determined from values obtained using
the Coulter Counter® Model S + STKR cell counting method. Viability
percentages were determined by Acridine Orange/Ethidium Bromide method.
RBC percentages were determined by hemo-cytometer count under a light
microscope. Results are obtained following the procedures recommended
by the manufacturer.
Table 1
Cell Percentages,
FICOLL™ HYPAQUE™ versus BD Vacutainer® CPT™ Tube Method (7)
| |
FICOLL™ HYPAQUE™
Method |
VACUTAINER®
CPT™ Tube Method |
| PARAMETER |
AVG% |
SD |
CV |
AVG% |
SD |
CV |
| Recovery |
68.2 |
10.322 |
15.122 |
63.0 |
11.722 |
18.622 |
| Purity |
| Total
Mononuclear Cells |
93.5 |
* |
* |
92.2 |
* |
* |
| Lymphocytes |
79.8 |
* |
* |
79.6 |
* |
* |
| Monocytes |
13.6 |
* |
* |
12.2 |
* |
* |
| Viability |
94.7 |
4.5 |
4.8 |
95.8 |
2.8 |
2.9 |
| RBC Contamination |
6.3 |
9.9 |
157.1 |
16.9 |
14.2 |
84.0 |
| Granulocyte
Contamination |
6.5 |
* |
* |
8.2 |
* |
* |
|
|