BD Vacutainer® PPT™ Plasma Preparation Tube
Procedure
- The BD PPT™Tube should be at room temperature (18-25°C)
and properly labeled for patient identification.
- Collect blood into the BD PPT™Tube using your institution's
recommended procedure for standard venipuncture technique and
sample collection.
- After collection of whole blood in the BD PPT™Tube, gently
invert the BD PPT™Tube 8 - 10 times.
- After mixing, store the BD PPT™Tube upright at room temperature
until centrifugation. Blood samples should be centrifuged within
two (2) hours of blood collection for best results. Centrifugation
of a sample at a period greater than 2 hours may require validation
by your institution or testing laboratory. Centrifuge BD PPT™Tube/blood
sample at room temperature (18-25°C) for a minimum of 10 minutes
at 1,100 RCF (Relative Centrifugal Force).
WARNING: Excessive centrifuge speed (over 10,000 RCF) may cause tube breakage and exposure to blood and possible injury. To calculate the correct centrifuge speed for a given RCF, use the following calculator. Simply enter two of the three values in the calculator and press the Calculate button.
- To obtain an undiluted plasma sample remove the BD Hemogard
Closure (See Instructions for Removal of BD Hemogard Closure
Section) and decant plasma into a separate vessel or aliquot plasma
into a separate vessel using a transfer pipette. NOTE: When using
a transfer pipette be sure NOT to disturb the barrier with the
tip of the pipette.
- Refer to the Kit or Assay Manufacture's Instructions for Sample Handling.
|
|
|
|