Recall notification - BD MAX System

Sep 24, 2018


BD has confirmed based on customer complaint investigations that a limited number of BD MAX Systems were manufactured with purchased optical filters that have shown a potential to produce a false positive result for Candida glabrata when processing samples positive for Trichomonas vagina/is during testing with the BD MAX Vaginal Panel. If the BD MAX Vaginal Panel was successfully verified/validated using the contrived ZeptoMetrix Vaginal Panel Control Panel, the risk of a clinical false positive C. glabrata result is low. The reported C. glabrata false positive results were all identified by customers when using the control material.

If you require further assistance, please contact BD Customer/Technical Support at 800-638-8663, Monday- Friday between 7:00am and 7:00pm (EST) in the United States.