Recall notification - BD MAX Enteric Bacterial Panel

Jan 14, 2019


Through internal testing, BD has observed a shift in PCR metrics for BD MAX Extended Enteric Bacterial Panel master mixes when used with sample buffer tubes from the BD MAX Enteric Bacterial Panel (EBP) lot numbers listed above. The result of this shift may cause false negative results with the BD MAX Extended Enteric Bacterial Panel when testing specifically for Enterotoxigenic Escherichia coli (ETEC) and Plesiomonas shigelloides targets of the BD MAX Extended Enteric Bacterial Panel assay.

If you are only using EBP, there is no impact for these targets with the lots listed above.

If you require further assistance, please contact BD Customer/Technical Support at 800-638-8663, Monday-Friday 7:00 am to 7:00 pm (EST).