BD Vacutainer® Venous collection BD Vacutainer® blood collection tubes

BD Vacutainer® EDTA Tubes

Overview

BD is the inventor of BD Vacutainer® blood collection tubes as well as the worldwide leader in blood tubes. We support our products with extensive clinical expertise and educational materials to help you achieve the best possible results.

true
Features and Benefits
Promotional Story
false
References
false
Specification

GTIN - each

00382903686612

1

GTIN - Case

50382903686617

1000

GTIN - Shelfpack

30382903686613

100


Quantity - Box

100

Quantity - Case

1000


Glass or Plastic

P

Tube Size

13x100

Draw Volume

6

Closure Type/Color

Conventional/lavender

Label Type

Paper

Additive/Concentration

K2EDTA 10.8 mg

GTIN

GTIN - each 00382903686612 1
GTIN - Case 50382903686617 1000
GTIN - Shelfpack 30382903686613 100

Packaging

Quantity - Box 100
Quantity - Case 1000

Product Basic Specification

Glass or Plastic P
Tube Size 13x100
Draw Volume 6
Closure Type/Color Conventional/lavender
Label Type Paper
Additive/Concentration K2EDTA 10.8 mg
References
false
Resources
References
false
Frequently Asked Questions

BD Vacutainer Plus Tubes are made of a special formulation of the plastic, polyethylene terephthalate (PET). This special formulation is extremely strong, dimensionally stable, and resistant to chemicals and breakage.

All non-gel separator blood collection tubes, including those that contain heparin, EDTA and non-gel serum tubes can be centrifuged at ≤1300 RCF for 10 minutes. The BD Vacutainer SST and PST gel tubes should be spun at room temperature at a speed of 1000 to 1300 RCF for 10 minutes in a swinging bucket centrifuge and 15 minutes in a fixed-angle centrifuge. The BD Vacutainer Barricor plasma blood collection tubes have a mechanical separator and must be spun at a minimum RCF of 1800 for 10 min in a swinging bucket centrifuge. BD Barricor tubes are optimally spun at an RCF of 4000 for 3 min in a swinging bucket centrifuge. BD Barricor tubes are not compatible with fixed angle centrifuges.

BD Vacutainer Plus plastic serum and SST tubes are coated with silicone and micronized silica particles to accelerate clotting. A silicone coating reduces adherence of red cells to tube walls. The silica coating can sometimes cause the inner tube wall to appear cloudy and/or filmy. This cloudy appearance does not make the tubes unacceptable for use.

Most BD Vacutainer blood collection tubes are packaged 100 tubes/box and 1,000 tubes/case (10 boxes).

No, BD does not claim that any of the Vacutainer blood collection tubes are endotoxin free.

The expiration date on the tube label is stated as a Year/Month/Day.

Local trash collectors should be consulted to see what the proper method of disposal is for unused, expired blood collection products.

Yes, BD Vacutainer blood collection tubes have a sterile interior. Tubes are sterilized by gamma radiation.

BD Vacutainer tubes are irradiated to achieve sterility. These tubes are sterile on the interior only. We cannot recommend re-sterilizing the tubes, i.e., ETO (ethylene oxide) or autoclaving, primarily due to pressure changes that take place during the re-sterilization cycle. Our tubes are under a specific negative pressure. During the re-sterilization cycle this negative pressure may be disrupted and therefore the tubes may not draw the proper blood volume.

BD does offer a sterile peel-apart pouch (reference number 366401) that contains a 10 mL glass red top serum tube and 7 mL glass lavender top EDTA tube.

BD Vacutainer Plus Tubes are made of a special formulation of the plastic, polyethylene terephthalate (PET). This special formulation is extremely strong, dimensionally stable, and resistant to chemicals and breakage.

All non-gel separator blood collection tubes, including those that contain heparin, EDTA and non-gel serum tubes can be centrifuged at ≤1300 RCF for 10 minutes. The BD Vacutainer SST and PST gel tubes should be spun at room temperature at a speed of 1000 to 1300 RCF for 10 minutes in a swinging bucket centrifuge and 15 minutes in a fixed-angle centrifuge. The BD Vacutainer Barricor plasma blood collection tubes have a mechanical separator and must be spun at a minimum RCF of 1800 for 10 min in a swinging bucket centrifuge. BD Barricor tubes are optimally spun at an RCF of 4000 for 3 min in a swinging bucket centrifuge. BD Barricor tubes are not compatible with fixed angle centrifuges.

BD Vacutainer Plus plastic serum and SST tubes are coated with silicone and micronized silica particles to accelerate clotting. A silicone coating reduces adherence of red cells to tube walls. The silica coating can sometimes cause the inner tube wall to appear cloudy and/or filmy. This cloudy appearance does not make the tubes unacceptable for use.

Most BD Vacutainer blood collection tubes are packaged 100 tubes/box and 1,000 tubes/case (10 boxes).

No, BD does not claim that any of the Vacutainer blood collection tubes are endotoxin free.

The expiration date on the tube label is stated as a Year/Month/Day.

Local trash collectors should be consulted to see what the proper method of disposal is for unused, expired blood collection products.

Yes, BD Vacutainer blood collection tubes have a sterile interior. Tubes are sterilized by gamma radiation.

BD Vacutainer tubes are irradiated to achieve sterility. These tubes are sterile on the interior only. We cannot recommend re-sterilizing the tubes, i.e., ETO (ethylene oxide) or autoclaving, primarily due to pressure changes that take place during the re-sterilization cycle. Our tubes are under a specific negative pressure. During the re-sterilization cycle this negative pressure may be disrupted and therefore the tubes may not draw the proper blood volume.

BD does offer a sterile peel-apart pouch (reference number 366401) that contains a 10 mL glass red top serum tube and 7 mL glass lavender top EDTA tube.

true
References
false
Related Products
RELATED PRODUCTS NOT AVAILABLE
References
false
Product Complaints
North American Regional Complaint Center
1-844-8BD-LIFE (1-844-823-5433)
Things to Consider

If you are a patient or end user, you can contact us yourself, or you may have your caregiver or your physician do that for you. To help us process your
information quickly and effectively, please contact our customer complaints
team.

To better facilitate our investigation, please include the following information in your reporting:


  • Product Name and/or Catalog Number
  • Lot Number or Serial Number
  • Any injuries and/or Harm?
  • What is the issue you experienced?
  • Is the actual sample or sample representative available? (If possible, please send affected sample)
  • Contact name and phone number
Product Recalls
Literature Icon
Recall Notifications
These recall notices provide current information on medical device recalls.
Learn more
References
false

Get a personalized experience by choosing your professional area

Medical Professional Lab Professional Pharmacist Other diagnostic area
true
Chat with us
Our live chat is available between the hours of 8.30am - 5.00pm EST, Monday - Friday
×